சுருக்கம்

Chemical characterization and degradation behavior of a polylactide surgical implant for the fixation of osteochondral fragments after activation with ultrasound energy

Benjamin Kienast, Nils Zander, Philip Henry, Bjoern Vahjen, Justus Gille, Christian J???????¼rgens, Arndt Peter Schulz


The purpose of this study was to demonstrate that the ultrasonic energy used to melt the tip of an absorbable polylactide Pin (SonicPin) does not have a significant effect on the chemical characterization and associated degradation properties of the material. The results indicate that the polylactide Pin after ultrasound is chemically equivalent to the SonicPin prior to ultrasound with regard to chemical structure and risk of leachable compounds. The evidence gathered from literature and testing suggests that any potential toxicological risk to the patient is not amplified by the application of ultrasonic energy to melt the tip of the device


மறுப்பு: இந்த சுருக்கமானது செயற்கை நுண்ணறிவு கருவிகளைப் பயன்படுத்தி மொழிபெயர்க்கப்பட்டது மற்றும் இன்னும் மதிப்பாய்வு செய்யப்படவில்லை அல்லது சரிபார்க்கப்படவில்லை

குறியிடப்பட்டது

  • CASS
  • கூகுள் ஸ்காலர்
  • ஜே கேட் திறக்கவும்
  • சீனாவின் தேசிய அறிவு உள்கட்டமைப்பு (CNKI)
  • CiteFactor
  • காஸ்மோஸ் IF
  • எலக்ட்ரானிக் ஜர்னல்ஸ் லைப்ரரி
  • டைரக்டரி ஆஃப் ரிசர்ச் ஜர்னல் இன்டெக்சிங் (DRJI)
  • ரகசிய தேடுபொறி ஆய்வகங்கள்
  • ICMJE

மேலும் பார்க்க

ஜர்னல் எச்-இண்டெக்ஸ்

Flyer